TB 9-6625-160-24
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range:
7 mV to 300 V
Fluke, Model 5720A (5720A) (p/o
MIS-35947); w/ amplifier, Fluke
Accuracy: 0. 5%
5725A/AR (5725A/AR)
Range:
20 Hz to 100 kHz
Accuracy: 5%
Flatness: 20 Hz to 100 kHz, < 6 V
Accuracy: 0.1%
Voltage range: 30 mV to 10 V ac
Hewlett Packard, Model 3458A
(3458A)
Accuracy:
0.15 %
Agilent, Model 54831M
(OS-303/G)
Sensitivity: 5 mV/div. min
100 kΩ
Range:
IET Labs, HARS-LX-9-0.001
(HARS-LX-9-0.001)
STANDARD
Accuracy: 0.25%
Fluke, Model 9640A-LPN
RF REFERENCE
(9640A-LPN)
SOURCE
Accuracy: 0.001% of setting
Range:
(-30dBm) 20 Hz to 100 kHz
Accuracy: 0.09 dB
Range:
-70 to +10 dBm 20 Hz to 300 kHz
Accuracy: 0.03 dB
1 Used
only for adjustments.
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the applicable sections before beginning
the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.