• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-86. Calibration Description.
Table B-92. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  34    35    36    37    38  39  40    41    42    43    44  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Ideal Industries, Model 61-746 - Continued
Table B-90. AC Current (≤1000 A) (w/50 Turn Coil).
Calibrator output
Ammeter calibrator
Test instrument
Applied
Range
Range
Limits (A ac)
(V ac)
(A ac)
(A ac)
Min
Max
1.08
20
600
521
559
1.08
20
600
521
559
1.08
20
600
514
566
Table B-91. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
400
-1.2105% (187.7)
+1.2105% (192.3)
1.9 k
4k
-1.2105% (1.877 k)
+1.2105% (1.923 k)
19 k
40 k
-1.2105% (18.77 k)
+1.2105% (19.23 k)
190 k
400 k
-1.2105% (187.7 k)
+1.2105% (192.3 k)
1.9 M
4M
-1.7368% (1.867 M)
+1.7368% (1.933 M)
19 M
40 M
-3.2632% (18.38 M)
+3.2632% (19.62 M)
B-26