TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Fluke, Model 902
Test instrument parameters
Performance specifications
DC VOLTAGE
Range: 0 to 600 V dc
Accuracy: (1.0% rdg + 5 dgts)
AC VOLTAGE
Range: 0 to 600 V ac
Accuracy: 50 to 60 Hz, (1.0% rdg + 5 dgts)
Range: 0 to 200 μA dc
DC CURRENT
Accuracy: (1.0% rdg + 5 dgts)
AC CURRENT
Range: 0 to 600 A ac
Accuracy: 50 to 60 Hz, (2.0% rdg + 5 dgts)
Range: 0 to 9999 Ω
Accuracy: (1.5% rdg + 5 dgts)
Range: 0 to 1000 μF
Accuracy: (1.9% rdg + 2 dgts)
TEMPERATURE
Range: -40 to 400 C
Accuracy: (1% + 0.8 C)
Table B-93. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
180
600
177.7
182.3
420
600
415.3
424.7
540
600
534.1
545.9
-540
600
-545.9
-534.1
Table B-94. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
540 @ 50 Hz
600
534.1
545.9
540 @ 60 Hz
600
534.1
545.9
Table B-95. DC Current (≤500 A)1.
Calibrator output
Test instrument
Limits (μA)
Applied
Range
(μA)
(μA)
Min
Max
180
200
177.7
182.3
Step 10 a (1), connect TI μAVΩ and COM terminals to calibrator output terminals and proceed to step 10 a (3).
1
Table B-96. AC Current (≤1000 A) (w/50 Turn Coil).
Calibrator output
Test instrument
Applied
Range
Range
Limits (A ac)
(V ac)
(A ac)
(A ac)
Min
Max
1.08 @ 50 Hz
20
600
528.7
551.3
1.08 @ 60 Hz
20
600
528.7
551.3
B-27