Table B-92. Calibration Description.
Table B-100. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  36    37    38    39    40  41  42    43    44    45    46  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Fluke, Model 902 Continued
Table B-97. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
999
-1.789% (186.6)
+1.789% (193.4)
1.9 k
9999
-1.789% (1866)
+1.789% (1934)
Table B-98. Capacitance.
Capacitance standard
Test instrument
Limits (μF)
Applied
Range
(μF)
(μF)
Min
Max
90
100
88.1
91.9
900
1000
881
919
Table B-99. Temperature.
Cool/Heat source
Test instrument
Limits ()
Applied
Range
()
Min
Max
-40 C
C
-43.5
-36.5
0C
C
-0.8
0.8
400 C
C
395.2
404.8
B-28