Table B-108. DC Current (=500 A).Calibrator output
Table B-114. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  40    41    42    43    44  45  46    47    48    49    50  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Meterman, Model AC75 Continued
Table B-113. Temperature.
Cool/Heat source
Test instrument
Limits ()
Applied
Range
()
Min
Max
-30 C
-40 to 400 C
-33.3 C
-26.7 C
360 C
-40 to 400 C
353.4 C
366.6 C
-30 F
-40 to 1000 F
-36.3 F
-23.7 F
900 F
-40 to 1000 F
903 F
915 F