Electro-Optical Gain Accuracy

Calibration Procedure For Night Vision Device (Nvd) Detector Standard, Model Dyn400 Page Navigation
  1    2    3    4  5
img
. . .
TB 9-6625-188-40
(7) Multimeter will indicate within 10.6% of value recorded in 7 k (5) above. If not,
perform b below.
b. Adjustments
(1) Gain access to TI adjustment by removing battery cover (fig. 2).
(2) Adjust TI to multimeter value recorded in 7 k (5) above.
9. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.