• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION I. IDENTIFICATION AND DESCRIPTION
SECTION II. EQUIPMENT REQUIREMENTS

Calibration Procedure For Spectrum Analyzer, If, Lf, And Rf Plug-In Units, Models 8552, 8553, 8554, 8555, And 8556 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
TB 9-6625-1914-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Model 8552( ) - continued
Calibrator output:
-30 dBm  0.3 dB
Amplitude
(All models)
30 MHz  300 kHz
Frequency
(Model 8552A)
30 MHz  3 kHz
(Model 8552B)
10% at 0.1 to 20 ms/div
Scan time accuracy
20% at 50 ms/div to 10 s/div
Model 8553 ( )
Frequency range
1 kHz to 110 MHz (Model 8553B has two ranges: 0 to 11 MHz
and 1 kHz to 110 MHz)
1 MHz of center frequency dial indication (Model 8553B,
Center frequency accuracy
0 to 11 MHz 200 kHz with FINE TUNE control
centered)
50  (models prefixed H01 and H02, 75  )
Input impedance
Noise sidebands
>70 dB below CW signal, 50 kHz or more away from CW
signal with 1 kHz IF bandwidth
0.5 dB from 1 kHz to 110 MHz for attenuator settings
Frequency response
>10 dB
Average noise level (sensitivity)
Model 8553L: -110 dBm-bandwidth = 10 kHz
Model 8553B: -100 dBm-bandwidth = 10 kHz
Local oscillator stability
Stabilized: <20 Hz p-p
Unstabilized: <1 kHz p-p
3% between 2 points except ranges between 50 kHz/div
Scan width accuracy
and 1 MHz/div 10%
Spurious responses
>70 dB below -40 dBm RF signal between 2 and 110 MHz
Model 8554( )
Frequency range
100 kHz to 1250 MHz (Model 8554L, 500 kHz to 1250 MHz
10%
Center frequency accuracy
50 (models prefixed H01 and H02, 75 )
Input impedance
Noise sidebands
>70 dB below CW signal, 50 kHz away from signal with 1 kHz
bandwidth (Model 8554L, >60 dB below CW signal, 20 kHz
away with 1 kHz bandwidth)
1 dB from 100 kHz to 1250 MHz (Model 8554L, 500 kHz to
Frequency response
1250 MHz)
Average noise level (sensitivity)
-102 dBm with 10 kHz IF bandwidth
Stabilized: 10 kHz/10 min
Local oscillator stability
Unstabilized: 50 kHz/10 min
Scan width accuracy
Error between 2 points on display is <10% of the indicated
separation
Spurious responses
>60 dB below a -40 dBm input RF signal (Model 8554B, >65 dB
below -40 dBm input)
Model 8555( )
15 MHz on fundamental mixing bands
Dial accuracy
Frequency range
0.01 to 18 GHz
Noise sidebands
At least 70 dB below CW signal level 30 kHz away from
signal using a 1 kHz IF bandwidth and 100 Hz video
filter
4


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.