Power Supply Check
Scan Width Accuracy

Calibration Procedure For Spectrum Analyzer, If, Lf, And Rf Plug-In Units, Models 8552, 8553, 8554, 8555, And 8556 Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . . . . . . . .
TB 9-6625-1914-24
SECTION IV
CALIBRATION PROCESS FOR
SPECTRUM ANALYZER RF PLUG-IN
HEWLETT-PACKARD, MODEL 8553( )
15. Preliminary Instructions
a. The instructions outlined in paragraphs 15 and 16 are preparatory to the calibration
process. Personnel should become familiar with the applicable section before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
NOTE
Unless otherwise specified, verify the result of each test and,
whenever the test requirement is not met, take corrective
action before continuing with the calibration.  Adjustments
required to calibrate the TI are included in this procedure.
Additional maintenance information is contained in the
manufacturer's manual for this TI.
NOTE
Unless otherwise specified, all controls and control settings
refer to the TI.
16. Equipment Setup
a. Mate TI with IF plug-in and install in display section.
b. Perform Section III, paragraph 7 b through 7 cc before beginning with paragraph 17.
17. Scan Width Accuracy
a. Performance Check
(1) Connect signal generator to RF INPUT.
(2) Position TI controls as listed in (a) through (h) below:
(a) RANGE MHz switch to 0-110 (not on all models).
(b) FREQUENCY control to 60 MHz.