TB 9-6625-1914-24
(g) SCAN TRIGGER switch to VIDEO.
(4) Adjust signal generator for output of 30 MHz at -30 dBm.
(5) Set signal generator controls as necessary for external modulation.
(6) Adjust function generator no. 1 controls to provide 90 percent modulation of
signal generator at 1 kHz.
(7) Fine tune signal generator frequency controls for best display.
(8) Adjust LINEAR SENSITIVITY (LOG REF LEVEL vernier) vernier control for
a convenient display height.
(9) Adjust DISPLAY ADJUST HORIZONTAL POSITION control to position first
modulation peak directly on -5 graticule line (left of centerline).
kHz, perform b below.
(11) Repeat technique of (4) through (10) above, using SCAN TIME PER DIVISION
switch settings listed in table 6. If function generator no. 1 indication is not within limits
specified, perform b below.
Test instrument
Function generator no. 1 frequency indications
SCAN TIME PER
DIVISION switch
Min
Max
settings
0.1 ms
9
kHz
11
kHz
0.2 ms
4.5
kHz
5.5
kHz
0.5 ms
1.8
kHz
2.2
kHz
1
ms
0.9
kHz
1.1
kHz
2
ms
450
Hz
550
Hz
5
ms
180
Hz
220
Hz
10
ms
90
Hz
110
Hz
20
ms
45
Hz
55
Hz
50
ms
16
Hz
24
Hz
0.1 s
8
Hz
12
Hz
0.2 s
4
Hz
6
Hz
b. Adjustments
(1) Connect TI and RF plug-in to extender and connect extender (supplied with TI)
to display section.
(2) Adjust function generator no. 1 frequency to 1 kHz and adjust A6R12 SCAN
TIME (fig. 3) or A6R15 SCAN TIME (fig. 2) to align 10th modulation peak with +4 graticule
line (R).
(3) Readjust A6R12 SCAN TIME or A6R15 SCAN TIME as necessary for best
compromise at all SCAN TIME PER DIVISION switch settings.
18