Table 6. Scan Time Accuracy
SECTION IV. CALIBRATION PROCESS FOR SPECTRUM ANALYZER RF PLUG-IN HEWLETT-PACKARD, MODEL 8553( )

Calibration Procedure For Spectrum Analyzer, If, Lf, And Rf Plug-In Units, Models 8552, 8553, 8554, 8555, And 8556 Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . . . . . . . . .
TB 9-6625-1914-24
13. Power Supply Check
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check
(1) Connect TI to extender and connect extender to display section.
(2) Connect multimeter between TI pin 11 of connector XA5 (fig. 3 or fig. 4)
(white/black/red/white) and chassis ground, using adapter and cable (B4 and B12).
(3) Adjust multimeter for dc measurement and slowly vary autotransformer output
between 105 and 125 V ac while observing multimeter. If multimeter does not remain
between +19.9 and +20.1 V dc, perform b (1) below.
(4) Adjust autotransformer for 115 V ac and measure ripple voltage. Multimeter
will indicate less than 0.5 mV ac.
(5) Connect multimeter between TI pin 8 of connector XA5 (fig. 3 or fig 4)
(white/black/violet wire) and chassis ground.
(6) Slowly vary autotransformer output between 105 and 125 V ac while measuring
dc voltage. If multimeter does not remain between -9.99 and -10.01 V ac, perform
b (2) below.
(7) Adjust autotransformer for 115 V ac and measure ripple voltage. Multimeter
will indicate less than 0.5 mV ac.
b. Adjustments
(1) Adjust A5R16 + 20 ADJ (fig. 3 or fig. 4) for a +20.00 V dc indication on multimeter.
(2) Adjust A5R32 -10 V ADJ (fig. 3 or fig. 4) for a -10.00 V dc indication on multimeter.
14. Final Procedure
a. De-energize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.