Table 2. Accessories Required
Figure 2. Thermistor mount/power sensor (10 to 40 GHz) - equipment setup.

Calibration Procedure For Power Sensors And Thermistor Mounts 10 Mhz To 40 Ghz (General) Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . . . .
TB 9-6625-1932-24
SECTION III
CALIBRATION PROCESS
(SECONDARY REFERENCE ONLY)
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire procedure before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 1.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
7. Equipment Setup
CAUTION
Do not twist the body of power sensor when connecting and
disconnecting. Twisting can cause major damage to the power
sensor circuit.
a. If Hewlett-Packard, Model 11528A adapter/cable is not available, fabricate
adapter/cable according to figure 1.
Figure 1. Power measurement system adapter/cable fabrication.
b. Select standard thermistor mount according to TI being calibrated.