Power Accuracy - Continued
Figure 4. VSWR (waveguide TIs, up to 18 GHz) - equipment setup.

Calibration Procedure For Power Sensors And Thermistor Mounts 10 Mhz To 40 Ghz (General) Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
.
TB 9-6625-1932-24
Figure 3. VSWR (coaxial TIs from 10 MHz to 18 GHz) - equipment setup.
(c) Add +0.712 to value recorded in (18) above (let value recorded in (18) above
be 10.741).
10.741
(18) above
+0.712
11.453
return loss
(21) Divide value recorded in (17) above by 2. Invert sign and add to value recorded
in (18) above. Record sum as return loss.
EXAMPLE B: If value of (17) above is +1.425
(a) Divide +1.425 by 2 = +0.712.
(b) Invert sign of +0.712 = -0.712.
(c) Add -0.712 to value recorded in (18) above (let value recorded in (18) above
be 10.741).
9