Table 1. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
ATTENUATOR (FIXED)
Range: 30 dB
Weinschel, Model 9918,
9918-30dB, and 9918-60dB
Frequency range: 10 MHz to 18 GHz
(9918)
Accuracy: 1.0 dB
Range: 60 dB
Frequency range: 10 MHz to 18 GHz
Accuracy: 1.5 dB
John Fluke, Model
Dc range: 0.1 V dc
8506A/CT (p/o MIS-35947)
Accuracy: 0.013% of indication +0.005% of FS
Dc range: 1 to 15 V dc
Accuracy: 0.013% of indication +0.002% of FS
Accuracy: 3%
Hewlett-Packard, Model
DIRECTIONAL
X752C (7923188)
COUPLER
Coupling: 10 dB
NO. 1
Coupling variation: 0.4 dB of nominal
DIRECTIONAL
Hewlett-Packard, Model
P752C (7923187)
COUPLER
Coupling: 10 dB
NO. 2
Coupling variation: 0.4 dB of nominal
PRD, Model K414-10-FS1
DIRECTIONAL
(7923152)
COUPLER
Coupling: 10 dB
NO. 31
Directivity: 40 dB
Coupling variation: 0.4 dB of nominal
PRD, Model A414-10-FS1
DIRECTIONAL
Frequency range: 26.5 to 40 GHz
(7923153)
COUPLER
Coupling: 10 dB
NO. 4
Directivity: 40 dB
Coupling variation: 0.4 dB of nominal
FREQUENCY
Weinschel, Model 1611
EXTENSION KIT NO. 1
(1611)
IF frequency: 700 MHz
Combined accuracy with receiver system: 0.03 dB/10dB
FREQUENCY
Weinschel, Model 1612
EXTENSION KIT NO. 2
(1612)
IF frequency: 1 GHz
Combined accuracy with receiver system: 0 .03 dB/10
dB
POWER BRIDGE
No substitute
NBS, Model Type II (Type
II)
POWER METER NO. 1
Furnished with TI
---
POWER METER NO. 2
Power range: -10 to +3 dBm
Hewlett-Packard, Model
E12-432A (MIS-30525)
See footnotes at end of table.