TB 9-6625-195-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set, AN/GSM-286; AN/GSM-287; or AN/GSM-440. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parentheses.
5. Accessories Required. The accessories required for this calibration are common
usage accessories issued as indicated in 4 above, and are not listed in this calibration
procedure.
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
Fluke, Model 5720A (5720A) (p/o
Range: 0 to 1000 V
MIS-35947); w/ amplifier, Fluke
5725A/AR (5725A/AR)
Accuracy: 0.0375%
Ac voltage
Range: 0 to 1000 V
Accuracy: 0.375%
Dc Current
Range: 0 to 10 A
Accuracy: 0.2%
Ac Current
Range: 0 to 10 A
Accuracy: 0.375%
Range:0 to 4 MΩ
Accuracy: 0.05%
Range: 40 MΩ to 400 MΩ
Accuracy: 0.25%
Agilent, Model 33250A (33250A)
FUNCTION
Range: 0 Hz to 100 kHz, 2 V rms
Accuracy: 0.0025%
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common