TB 9-6625-1959-24
(8) Adjust FREQUENCY VARIABLE control for a display of the next frequency
listed in table 4 on FREQUENCY MHz indicator.
(9) Repeat steps (3) (d) through (7) above:
(10) Repeat steps (8) and (9) above, for the remaining TI FREQUENCY VARIABLE
Test instrument
Spectrum analyzer
FREQUENCY
RANGE
FREQUENCY
(MHz)
VARIABLE
Indications
switch
control
Span
(dB)
minimum ( )
settings
settings
setting
frequency
2d
100-250
250
2 MHz
500
MHz
-35
3d
2 MHz
750
MHz
-40
2d
175
2 MHz
350
MHz
-35
3d
2 MHz
525
MHz
-40
2d
100
2 MHz
200
MHz
-35
3d
2 MHz
300
MHz
-40
2d
50-100
50
2 MHz
100
MHz
-35
3d
2 MHz
150
MHz
-40
2d
75
2 MHz
150
MHz
-35
3d
2 MHz
225
MHz
-40
2d
100
2 MHz
200
MHz
-35
3d
2 MHz
300
MHz
-40
2d
25-50
50
2 MHz
100
MHz
-35
3d
2 MHz
150
MHz
-40
2d
37.5
2 MHz
75
MHz
-35
3d
2 MHz
112.5 MHz
-40
2d
25
2 MHz
50
MHz
-35
3d
2 MHz
75
MHz
-40
2d
10-25
10
2 MHz
20
MHz
-35
3d
2 MHz
30
MHz
-40
2d
17.5
2 MHz
35
MHz
-35
3d
2 MHz
52.5
MHz
-40
2d
25
2 MHz
50
MHz
-35
3d
2 MHz
75
MHz
-40
2d
5-10
10
2 MHz
20
MHz
-35
3d
2 MHz
30
MHz
-40
2d
7.5
2 MHz
15
MHz
-35
3d
2 MHz
22.5
MHz
-40
2d
5
2 MHz
10
MHz
-35
3d
2 MHz
15
MHz
-40
2d
2.5-5
2.5
2 MHz
5
MHz
-35
3d
2 MHz
7.5
MHz
-40
2d
3.75
2 MHz
7.5
MHz
-35
3d
2 MHz
11.25 MHz
-40
2d
5
2 MHz
10
MHz
-35
8