Table 6. Flatness - Continued

Calibration Procedure For Leveled Sine Wave Generator Type Sg-503 Page Navigation
  7    8    9    10    11  12
img
. . . .
TB 9-6625-1959-24
(2) Connect multimeter HI to +5.2 VOLT TP (fig. 2). Multimeter will indicate
between 5.0 and 5.4 V dc.
b. A d j u s t m e n t s . A d j u s t R 6 9 4 - 2 2 V O L T A D J ( f i g . 2 ) f o r a - 2 2 V i n d i c a t i o n
o n multimeter.
13. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
13/14 Blank