TB 9-6625-1959-35
(c) Marker, Delta.
(e) Peak Search.
(5) The spectrum analyzer ∆Mkr will indicate less than or equal to the minimum
limit listed in table 3. If spectrum analyzer is not less than or equal to the minimum limit
Table 3. Output Buffer Current and Harmonic Suppression
Spectrum analyzer
Test
Indication
Instrument
(dB)
minimum (≤)
frequency Harmonic
(MHz)
100 MHz
200 MHz
2nd
-35
100 MHz
300 MHz
3rd
-40
(7) Adjust OUTPUT AMPLITUDE VOLTS P-P control to 3.0.
(8) Repeat (3) through (6) above.
(9) Adjust OUTPUT AMPLITUDE VOLTS P-P control to 0.5.
(10) Repeat (3) through (6) above.
b. Adjustments.
Adjust CURRENT ADJ
R175 (fig. 2) until third
harmonic is at
least
40 dB down from reference while assuring that second harmonic amplitude remains 35 dB
down from reference. If second harmonic is equal to or greater than 35 dB down, do not
adjust CURRENT ADJ R175 (fig. 2) to reduce second harmonic at expense of increasing the
third (R).
Figure 2. Main
circuit board adjustment locations -
right side.
6
CHANGE 1