TB 9-6625-1959-35
(17) Disconnect precision cable from 50Ω feedthrough termination and connect to
measuring receiver power sensor using necessary adapter.
(19) Set TI controls to the first FREQUENCY VARIABLE control setting listed in
(20) Set the measuring receiver to the first FREQUENCY VARIABLE control
setting listed in table 6 using the numeric key pad and the MHz keys.
NOTE
Do not take the measuring receiver out of the ratio mode
(21)
Measuring receiver will indicate between the minimum and maximum values
listed in table 6 for the selected frequency
(22)
Repeat (19) through (21) above for each remaining frequency listed in table 6.
Measuring receiver
Test instrument
indication
FREQUENCY
RANGE
FREQUENCY
(MHz)
VARIABLE
switch
control
settings
settings
Min
Max
.25-.5
.25
98.5%
101.5%
.375
98.5%
101.5%
.50
98.5%
101.5%
.5-1.0
1.0
98.5%
101.5%
.75
98.5%
101.5%
.50
98.5%
101.5%
1-2.5
1.0
98.5%
101.5%
1.25
98.5%
101.5%
1.50
98.5%
101.5%
1.75
98.5%
101.5%
2.00
98.5%
101.5%
2.25
98.5%
101.5%
2.50
98.5%
101.5%
2.5-5.0
5.00
98.5%
101.5%
4.50
98.5%
101.5%
4.00
98.5%
101.5%
3.50
98.5%
101.5%
3.00
98.5%
101.5%
2.50
98.5%
101.5%
5.0-10
5.0
98.5%
101.5%
6.0
98.5%
101.5%
7.0
98.5%
101.5%
8.0
98.5%
101.5%
9.0
98.5%
101.5%
10
98.5%
101.5%
11