TB 9-6625-1959-35
(8) Adjust FREQUENCY VARIABLE control for a display of the next frequency
listed in table 4 on FREQUENCY MHz indicator.
(9) Repeat steps (3)(d) through (7).
(10) Repeat steps (8) and (9) for the remaining TI FREQUENCY VARIABLE
Test instrument
Spectrum analyzer
FREQUENCY
RANGE
FREQUENCY
(MHz)
VARIABLE
Indications
switch
control
Span
(dB)
minimum (≤)
settings
settings
setting
frequency
100-250
250
2 MHz
500
MHz
2nd
-35
2 MHz
750
MHz
3rd
-40
175
2 MHz
350
MHz
2nd
-35
2 MHz
525
MHz
3rd
-40
100
2 MHz
200
MHz
2nd
-35
2 MHz
300
MHz
3rd
-40
50-100
50
2 MHz
100
MHz
2nd
-35
2 MHz
150
MHz
3rd
-40
75
2 MHz
150
MHz
2nd
-35
2 MHz
225
MHz
3rd
-40
100
2 MHz
200
MHz
2nd
-35
2 MHz
300
MHz
3rd
-40
25-50
50
2 MHz
100
MHz
2nd
-35
2 MHz
150
MHz
3rd
-40
37.5
2 MHz
75
MHz
2nd
-35
2 MHz
112.5 MHz
3rd
-40
25
2 MHz
50
MHz
2nd
-35
2 MHz
75
MHz
3rd
-40
10-25
10
2 MHz
20
MHz
2nd
-35
2 MHz
30
MHz
3rd
-40
17.5
2 MHz
35
MHz
2nd
-35
2 MHz
52.5
MHz
3rd
-40
25
2 MHz
50
MHz
2nd
-35
2 MHz
75
MHz
3rd
-40
5-10
10
2 MHz
20
MHz
2nd
-35
2 MHz
30
MHz
3rd
-40
7.5
2 MHz
15
MHz
2nd
-35
2 MHz
22.5
MHz
3rd
-40
5
2 MHz
10
MHz
2nd
-35
2 MHz
15
MHz
3rd
-40
2.5-5
2.5
2 MHz
5
MHz
2nd
-35
2 MHz
7.5
MHz
3rd
-40
3.75
2 MHz
7.5
MHz
2nd
-35
2 MHz
11.25 MHz
3rd
-40
5
2 MHz
10
MHz
2nd
-35
PIN: 047612-001