Dc Volts
Resistance

Calibration Procedure For Digital Multimeter Models 7D13 And 7D13A Page Navigation
  1    2    3    4  5  6    7    8  
img
.
TB 9-6625-1962-24
NOTE
Changing the setting of R1003 will affect the calibration of DC
current and resistance measurement modes on 7D13A.
Figure 1. Left side view - adjustments and test points.
6