Phase Lock
SECTION V. CALIBRATION PROCESS FOR MODEL 3304A

Calibration Procedure For Function Generator Model 3300A With Aux Unit Plug-In, Model 3301A, Trigger/Phase Plug-In, Model 3302A; Sweep/Offset Plug-In, Model 3304A; And Sweep Plug-In Model 3305A Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . .
TB 9-6625-1966-24
22. Phase Dial and Distortion
a. Performance Check
(1) Adjust function generator FREQUENCY for 1 kHz and set function switch to SINE.
(2) Set TI + INPUT PHASE -switch to + (positive).
(3) Adjust calibrator frequency to 1 kHz and for a 3.535 V output.
(4) Set oscilloscope vertical MODE switch to ALTERNATE, adjust horizontal
TIME/DIV switch for several cycles of display, and set TRIGGER SOURCE switch to
EXT.
(5) While observing oscilloscope, obtain phase lock by adjusting TI PHASE control
until two displays on oscilloscope are in phase. If TI PHASE dial does not indicate
0 20 degrees, perform b (1) below.
(6) Set TI + INPUT PHASE - switch to - (negative).  Repeat (5) above.
If TI
PHASE dial does not indicate 180  10 degrees, perform b (2) below.
(7) Repeat (2) through (6) above for frequencies of 100 Hz and 10 kHz, respectively.
Figure 5. Model 3302A - top view.
19