Equipment Setup
Trigger Amplitude

Calibration Procedure For Function Generator Model 3300A With Aux Unit Plug-In, Model 3301A, Trigger/Phase Plug-In, Model 3302A; Sweep/Offset Plug-In, Model 3304A; And Sweep Plug-In Model 3305A Page Navigation
  9    10    11    12    13  14  15    16    17    18    19  
img
. . .
TB 9-6625-1966-24
Figure 3. Single and multiple cycle - equipment setup.
(5) Adjust function generator frequency for 100 kHz.
(6) Adjust oscilloscope horizontal TIME/DIV switch for a convenient viewing display.
(7) Adjust TI START/STOP PHASE control until function generator output cycle
begins at zero-degree phase on positive slope of calibrator waveform, as shown in figure 4.
Figure 4. Single cycle phase relationship.