Sine Wave Response
Equipment Setup

Calibration Procedure For Function Generator Model 3300A With Aux Unit Plug-In, Model 3301A, Trigger/Phase Plug-In, Model 3302A; Sweep/Offset Plug-In, Model 3304A; And Sweep Plug-In Model 3305A Page Navigation
  7    8    9    10    11  12  13    14    15    16    17  
img
. . . . . . . . . . . .
TB 9-6625-1966-24
Table 5. Sine Wave Response Continued
Test instrument
Multimeter indications (V Ac)
RANGE
FREQUENCY
Min
Max
switch settings
dial settings
X1K
5
4.95
5.05
X1K
10
4.95
5.05
X10K
1
4.85
5.15
X10K
5
4.85
5.15
X10K
10
4.85
5.15
b. Adjustments. No adjustments can be made.
15. Power Supply
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check. Connect multimeter between test points and chassis ground
as listed in table 6 and shown in figure 2. If multimeter does not indicate within limits
specified, perform appropriate adjustments listed in table 6.
Table 6. Power Supply
Test instrument
Multimeter
Adjustments
(fig. 2)1
test points
indications (V dc)
(fig. 2)
Min
Max
Adjustments
Indications
A12C2 (A2C2)
-26.48
-26.52
A12R20 (A2R20)
-26.5
A12C1 (A2C1)
+26.48
+26.52
A12R7 (A2R7)
+26.5
A12C4 (A2C4)
-19.99
-20.01
A12R26 (R3)
-20.0
A12C3 (A2C3)
+19.99
+20.01
A12R25 (R2)
+20.0
1All
power supply adjustments interact; readjust as necessary for best compromise.
b. Adjustments. No further adjustments can be made.
16. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.
SECTION IV
CALIBRATION PROCESS FOR MODEL 3302A
17. Preliminary Instructions
a. The instructions outlined in paragraphs 17 and 18 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.