SECTION III. CALIBRATION PROCESS
Table 3. Dial Accuracy

Calibration Procedure For Function Generator Sg-1133U Model 3312A Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . .
TB-9-6625-1968-24
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel  fail  to  observe  safety  precautions.  REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Connect TI to a 115 V ac power source.
b. Position TI controls as listed in (1) through (9) below:
(1) TRIGGER INT/EXT switch (rear panel) to INT.
OFFSET control to CAL.
(2)
AMPLITUDE switch to .01 and AMPLITUDE VERNIER control fully ccw.
(3)
SYM control to CAL.
(4)
TRIGGER PHASE control to FREE RUN.
(5)
All MODULATION pushbuttons released to out.
(6)
MODULATION RANGE Hz switch 0.
(7)
MODULATION SYM control to CAL.
(8)
LINE OFF/ON pushbutton pressed to ON and allow 1 hour for warm-up.
(9)
8. Frequency Accuracy
a. Performance Check
f
(1) Connect TI SYNC to frequency counter CHANNEL A input using 50
eedthrough termination.
(2) Position TI controls as listed in (a) through (d) below:
FUNCTION sine wave pushbutton pressed.
(a)
RANGE Hz .1 pushbutton pressed.
(b)
FREQUENCY dial to 1.
(c)
AMPLITUDE switch to 1 and AMPLITUDE VERNIER control fully cw.
(d)
(3) If frequency counter does not indicate between 6060 ms and 28571 ms, perform
8 b below.
(4) Repeat technique of (2) and (3) above using settings and indications listed in
table 3.
4