Figure 2. 1 mVAC equipment setup.
Equipment Setup

Calibration Procedure For Ac Calibrator, Model 5200A And Power Amplifiers, Models 5215A And 5205A Page Navigation
  9    10    11    12    13  14  15    16    17    18    19  
img
. . . . . . . . . . . .
TB 9-6625-1976-24
(5) Ensure that VOLTAGE RANGE switch is set to 10 V and VOLTAGE dials to
1.00000 V (1 MHz).
(6) Adjust A8R89 for the same error as recorded in step 4 of table 6 (10 V, 1 MHz) 0.2% (R).
(7) Repeat technique of steps 4 and 5 of table 6 and b (5) and (6) above until they
can be performed without adjustments.
(8) Adjust A8R91 for best null indication on multimeter (R).
(9) Adjust A6C3 for a 0.033 ( 5 digits) indication on multimeter (R).
12. Power Supply
NOTE
Do not perform power supply checks if all other parameters are
within tolerance.
a. Performance Check
(1) Connect multimeter to REGULATOR A5A2TP8 (HI) and A5A2TP5 (LO) (fig. 1)
(2) Set MODE switch to OPER. If the multimeter does not indicate between 14.99
and 15.01 V dc, perform b (1) below.
(3) Move multimeter (HI) lead from REGULATOR A5A2TP8 to A7TP3 (fig. 1). If
multimeter does not indicate between 185 to 200 V dc, perform b (2) below.
(4) Set MODE switch to STDBY.
b. Adjustments
0.01 V dc indication on
(1) Adjust REGULATOR A5A2R32 (fig. 1) for a 15
multimeter (R).
(2) Adjust REGULATOR A5A2R32 (fig. 1) for a minimum indication on multimeter
of 185 V up to 220 V dc (R).
13. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
SECTION IV
CALIBRATION PROCESS FOR
PRECISION POWER AMPLIFIERS, FLUKE, MODELS 5215A AND 5205A
14. Preliminary Instructions
a. The instructions outlined in paragraphs 14 and 15 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.