Table 1. Calibration Description -- Continued
Equipment Setup

Calibration Procedure For Counters Types Dc 501, Dc 501Opt01, Dc 503, Dc 503Opt01, Dc 503A, Dc 503Aopt01, Dc 504, Dc 504A, Dc 508, Dc 508Opt01, Dc 508Aopt01 And Dc 509 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . . . . . . . . . .
TB 9-6625-1982-24
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.  Where the four-to-one ratio
cannot be met, the accuracy of the equipment selected is shown in parentheses.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: mainframe power module (furnished with TI) Tektronix TM506 (MIS-30526/1
Type I) or TM515 (MIS-30526/1 Type II) and extender Tektronix 067-0645-02.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
1 part in 109
FREQUENCY
Range:
Tracor, Model 527E (527E)
DIFFERENCE METER
Resolution: 1 part in 1010
Agilent, Model 33250A (33250A)
FUNCTION/ARBITRARY
Range:
10 Hz to 10 MHz
GENERATOR
Amplitude: 0 to 50 mV rms
Accuracy:  3%
MULTIMETER
Range:
0 15 V
Fluke, Model 8846A
Accuracy:
1.25%
Aeroflex, Model 2023B (2023B)
SIGNAL
Range:
10 MHz to 1.3 GHz
GENERATOR
Amplitude:
0 to 175 mV rms
Accuracy:
3%
TIME/FREQUENCY
Range:
1 MHz or 10 MHz
Datum, Model ET6000-75 (13589305)
WORKSTATION
1 part in 109
Accuracy:
SECTION III
CALIBRATION PROCESS FOR TEKTRONIX, TYPE DC501, AND DC501OPT01
6. Preliminary instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire section before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
6