TB 9-6625-1982-24
CALIBRATION PROCESS FOR TEKTRONIX, TYPES DC503, DC503OPT01
12. Preliminary Instructions
a. The instructions outlined in paragraphs 12 and 13 are preparatory to the calibration
process. Personnel should become familiar with the entire section before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in paragraph 15 are not within tolerance, perform the
power supply check and adjustments in paragraph 16. After adjustments are made repeat
paragraphs 14 and 15. Do not perform power supply check if all other parameters are
within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance checks where applicable.
CAUTION
To avoid damage to the plug-in circuitry, power must be OFF
on the mainframe power module before connecting the TI.
a. With power OFF, connect TI to mainframe power module using extender.
b. Pull mainframe power module ON-OFF PWR switch to ON and allow at least 30
minutes (60 minutes if stored in high humidity environment) for equipment to warm-up
and stabilize.
c. Remove left or right side covers of TI as necessary to make connections or
adjustments. Replace cover(s) after completing adjustments.