TB 9-6625-1995-35
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
OSCILLOSCOPE
Time markers:
Fluke, Model 5820A-5C-GHZ
CALIBRATOR
Range: .2 s/div to 2 ns/div
(5820A-5C-GHZ)
Accuracy: 0.5%
VERTICAL
Compatible with TI and supplied by user
Tektronix, Type 7A Series (7A
AMPLIFIER
Series)
1Use
recorded indications.
6. Preliminary
Instructions
a. The
instructions outlined in this
section are preparatory to the
calibration process.
Personnel should become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this bulletin. Additional
maintenance information is contained in the manufacturer's manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove protective covers from TI only when necessary to make adjustments.
Replace protective covers after completing the adjustments.
b. Install TI into
oscilloscope horizontal compartment and
vertical amplifier into
oscilloscope left vertical compartment.
c. Connect oscilloscope to a
115 V ac
source and
energize. Allow at
least 20
minutes for
warm-up.