High RF Voltage Range
IF Output Frequency Accuracy

Calibration Procedure For Signal Generator An/Urm-103 (Sg-297/U) Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
. . . .
TB 9-6625-1997-24
Table 6. High RF Voltage Range Accuracy
Test Instrument
Measuring Receiver
RF TUNING
RF OUTPUT
BAND
control for
SWITCH
switch setting
frequency
Min
Max
setting
(KUV)
(MHz)
(mV)
(mV)
B
30
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
A
20
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
C
45
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
D
65
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
(3) Repeat technique of (2) above for switch settings listed in table 6. If measuring
receiver does not indicate within limits specified, perform b below.
b. Adjustments
(1) Set RF OUTPUT switch to 250 KUV and adjust LO-HI RF SET TO LINE
control for a 250 mV indication on measuring receiver.
NOTE
When TI protective cover is removed to make adjustments, do
not change setting of 115 230 LINE VOLTAGE SELECTOR
switch (located on bottom front center).
(2) Adjust R1 ((HI-RF METER CAL) located on lower right side) for a 0.55 or a red-
line indication on IF UV RF SET TO LINE meter (R).
11. IF Output Level
a. Performance Check
(1) Connect measuring receiver to impedance matching adapter No. 1 located on TI
IF output.
(2) Position controls as listed in (a) through (c) below: