TB 9-6625-2011-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin
provides instructions for the
calibration of Pulse Generator, SG-1105/U and Hewlett-Packard, Model 8013B. The
manufacturer's manual was used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test
instrument) throughout this bulletin.
a. Model Variations . None.
b. Time and Technique . The time required for this calibration is approximately 3
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports.
a.
Forms, records, and
reports required for
calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Repetition rate
Range: 1 Hz to 50 MHz
Pulse delay
<35 ns to 1 s
Pulse width
<10 ns to 1 s
Pulse shape
Rise and falltime
<3.5 ns
Preshoot
<5%
Overshoot
<5%
Ringing
<5%
Double pulse
Equal pulses when spaced at least 20 ns
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.