TB 9-6625-2011-35
a. Performance Check
(1) Connect multimeter between +17 V TP (fig. 1) and ground. If multimeter does
not indicate between 16.9 and 17.1 V dc, perform b(1) below.
(2) Connect multimeter between -17 TP (fig. 1) and ground. If multimeter does
not indicate between -16.9 and -17.1 V dc, perform b(2) below.
b. Adjustments
(1) Adjust A6R102 (fig. 1) for a 17.00 V dc indication on multimeter (R).
(2) Adjust A6R106 (fig. 1) for a 17.00 V dc indication on multimeter (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9