TB 9-6625-2024-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
DUAL TIME BASE TEKTRONIX TYPES 7B53AN OPT 5 AND OPT 11S, 7B53A,
AND TD-1159/U (Continued)
Delayed sweep time
Range:
0.05 Ps/div to 0.5 s/div
(over center eight divisions)
Accuracy: r3% unmag, 3.5% mag from 50 ms/div to
0.5 Ps/div
r4% unmag, 4.5% mag from 0.5 s/div to
0.1 s/div and 0.2 Ps/div to 0.05 Ps/div
Variable time delay
Range:
0 to 10 times the TIME/DIV OR DLY TIME
Differential delay time measurement
control settings from 5 s/div to 1Ps/div
Accuracy: 5 s/div to 1 s/div, within 1.4% of
measurement plus 0.3% of FS1
0.5 s/div to 1 Ps/div, within 0.7%
of measurement plus 0.3% of FS1
1Full
scale is 10 times the TIME/DIV OR DLY TIME setting.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for the calibration are common
usage accessories, issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Extender, Tektronix Type 067-0589-00, Calibration Adapter, Tektronix Type
067-0587-01.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
DUAL TIME BASE1
Must be compatible with required
Tektronix, Type 7B53A or 7B53AN,
7B53AOPT5 or 7B53AN11S or TD 1085/U
or TD-1159/U
Range: -3049 to +140 V dc
Agilent, Model 3458A (3458A)
w/high voltage probe (3490AOPT060)
Accuracy: r0.025%
Must be compatible with required
Tektronix, Type 7603 or OS-245 (P)/U
dual time base and vertical
amplifier
See footnote at end of table.