SECTION II. EQUIPMENT REQUIREMENTS
Equipment Setup

Calibration Procedure For Oscilloscope Os-245(P)/U, And Types 7603 And 7603N Opt 11S, Vertical Amplifier Am-6565/U, And Types 7A15A And 7A15An Opt 11, And Dual Time Base Td-1085/U And Td-1159/U And Types 7B53A, 7B53An Opt 5 And Opt 11S Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . . .
TB 9-6625-2024-24
Table 2. Minimum Specifications of Equipment Required - Continued
Minimum use
Manufacturer and model
Common name
specifications
(part number)
OSCILLOSCOPE
Time markers:
Fluke, Model 5820A-5C-GHZ
CALIBRATOR
Range: 10 ns to 5 s
(5820A-5C-GHZ)
Accuracy: r0.75%, 10 to 20ns
r0.5%, 50 ns to 0.1 s
0.75%, 0.2 to 0.5 s
Voltage amplitude:
Range: 20 mV to 50 V at 1 kHz
Accuracy:  r0.5%
Pulses:
Risetime:  1 ns or less
VERTICAL
Must be compatible with required
Tektronix, Type 7A15A, or AM6565U
1
AMPLIFIER
dual time base and oscilloscope
1Normally
supplied as a system.
SECTION III
CALIBRATION PROCESS FOR OSCILLOSCOPE OS-245(P)/U AND TEKTRONIX
TYPES 7603 AND 7603N OPT 115
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by
common name and item identification number as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manuals and TM 11-6625-
2658-14 for these TIs.
d. When indications specified in paragraphs 8 through 12 are not within tolerance,
perform the power supply check prior to making adjustments.  After adjustments are
made, repeat paragraphs 8 through 12. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
5