TB 9-6625-2049-24
(11) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.598 and
3.176V ac.
b. Adjustments. No adjustments can be made.
17. Dc Offset
a. Performance Check
f
(1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50
eedthrough termination.
(2) Press SIGNAL key off (option 002 not lit).
(3) Press presently active FUNCTION key to remove ac output and activate
ENTRY DC OFFSET key (lit).
(4) Enter 5 VOLT using DATA keys and press AMPTD CAL key.
(5) Multimeter will indicate between 4.98 and 5.02 V dc.
(6) Enter -5 VOLT using DATA keys. Multimeter will indicate between -4.98
and -5.02 V dc.
(7) Repeat technique of (6) above using TI settings and multimeter indications
listed in table 5.
Table 5. Dc Offset
Test instrument
Multimeter indications
DC OFFSET
(V dc)
settings
Min
Max
1.499  VOLT
1.493
1.505
-1.499  VOLT
-1.505
-1.493
499.9
mV
0.4979
0.5019
-499.9
mV
-0.5019
-0.4979
149.9
mV
0.1493
0.1505
-149.9
mV
-0.1505
-0.1493
49.99
mV
0.04979
0.05019
-49.99
mV
-0.05019
-0.04979
14.99
mV
0.01493
0.01505
-14.99
mV
-0.01505
-0.01493
4.999  mV
0.004979
0.005019
-4.999  mV
-0.005019
-0.004979
1.499  mV
0.001479
0.001519
-1.499  mV
-0.001519
-0.001479
NOTE
Perform (8) through (11) below for option 002 only.
(8) Remove 50
feedthrough termination from equipment setup.
(9) Press SIGNAL key on.
(10) Enter 20 VOLT using DATA keys. Multimeter will indicate between 19.775
and 20.225 V dc.
18