TB 9-6625-2049-24
(10) Press FUNCTION triangle key and enter 9.9 kHz using DATA keys.
(11) Multimeter will indicate between -4.44 and -4.56 V dc.
(12) Press FUNCTION positive ramp key. Multimeter will indicate between -4.38
and -4.62 V dc.
b. Adjustments. No adjustments can be made.
19. High Voltage Output Option 002 only
a. Performance Check
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below:
FUNCTION sine wave.
(a)
ENTRY FREQ.
(b)
2 kHz.
(c)
SIGNAL on.
(d)
ENTRY DC OFFSET.
(e)
0 VOLT.
(f)
ENTRY AMPTD.
(g)
14.14 V RMS.
(h)
AMPTD CAL.
(i)
(3) Multimeter will indicate between 13.86 and 14.42 V ac.
(4) Press keys and enter values using DATA keys as listed in (a) through (e) below:
ENTRY FREQ.
(a)
100 kHz.
(b)
ENTRY AMPTD.
(c)
10 V RMS.
(d)
AMPTD CAL.
(e)
(5) Multimeter will indicate between 9 and 11 V ac.
(6) Press keys and enter values using DATA keys as listed in (a) through (f) below:
FUNCTION square wave.
(a)
ENTRY FREQ.
(b)
2 kHz.
(c)
ENTRY AMPTD.
(d)
20 V RMS.
(e)
AMPTD CAL.
(f)
(7) Multimeter will indicate between 19.6 and 20.4 V ac.
(8) Press FUNCTION triangle key and enter 11.5 V RMS using DATA keys.
(9) Press AMPTD CAL key. Multimeter will indicate between 11.27 and 11.73 V ac.
(10) Press FUNCTION positive ramp key and repeat (9) above.
b. Adjustments. No adjustments can be made.