Figure 5. Power supply assembly A2 test instrument top view.

Calibration Procedure For Synthesizer/Function Generator Models 3325A And 3325B Page Navigation
  17    18    19    20    21  22
img
. . .
TB 9-6625-2049-24
(11) Press AMPTD CAL key. Adjust R22 (fig. 5) until multimeter indicates 10.000 V dc.
(12) Enter -5 VOLT using DATA keys.
(13) Multimeter will indicate between -9.985 and -10.015 V dc.
b. Adjustment. No further adjustments can be made.
21. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.