TB 9-6625-2050-24
Test instrument parameters
Performance specifications
Range: 10 W to 100 mW
Meter accuracy:
1% FS on all ranges
Models 435A and 435B
Accuracy:
0.5% FS watt mode:
Models 436A and
Accuracy:
0.02 dB, 0.001 dB/oC, dB mode
436OPT9/22
0.02 dB, 0.001 dB/oC, dB [REL] mode
Zero carryover:
0.5% FS (zeroed on most sensitive scale)
Models 435A, 435BOPT001,
and 435BE21
0.2% FS (zeroed on most sensitive scale
Models 436A and
436AOPT9/22
1 mW at 50 MHz 0.7%
(Models 435A,
435BOPT001, and
435BE21, 436A and
436AOPT9/22)
50 MHz, 1%
(Models 436A and
436AOPT9/22)
Calibration factor
Range: 85 to 100% in 1% steps
6 mV
(Models 435A,
Accuracy:
435BOPT001, 435BE21,
436A, and 436AOPT9/22)
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. When the four-to-one ratio cannot
be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.