TB 9-6625-2050-35
Test instrument parameters
Performance specifications
Meter accuracy:
Range: 10 W to 100 mW
Models 435A and 435B
Accuracy: 1% FS on all ranges
Models 436A and
0.5% FS watt mode:
Accuracy:
436OPT9/22
0.02 dB, 0.001 dB/oC, dB mode
0.02 dB, 0.001 dB/oC, dB [REL] mode
Zero carryover:
Models 435A, 435BOPT001,
0.5% FS (zeroed on most sensitive scale)
and 435BE21
Models 436A and
0.2% FS (zeroed on most sensitive scale
436AOPT9/22
1 mW at 50 MHz 0.7%
(Models 435A,
435BOPT001, and
435BE21, 436A and
436AOPT9/22)
50 MHz, 1%
(Models 436A and
436AOPT9/22)
Calibration factor
Range: 85 to 100% in 1% steps
(Models 435A,
Accuracy: 6 mV
435BOPT001, 435BE21,
436A, and 436AOPT9/22)
EQUIPMENT REQUIREMENTS
4. Equipment
Required.
Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-287, or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. When the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories listed in
table 3 are issued as
indicated in
paragraph 4 above and
are used in this
calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.