TB 9-6625-2050-35
PRELIMINARY OPERATIONS
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUTS(S) to minimum after each step within the
performance check where applicable.
b. Connect autotransformer to a
115 V ac
source and
adjust for a
115 V ac
output.
c. Set LINE switch to ON and allow 30 minutes for TI to warm-up and stabilize.
CAUTION
To avoid damage to meter, set range calibrator FUNCTION
switch to STANDBY while changing RANGE switch settings
NOTE
If an out-of-tolerance indication is obtained during any
performance check, re-zero TI as follows: Set range calibrator
FUNCTION switch to STANDBY, then press TI ZERO
(model 435A) switch and wait for meter zero indication (model
435A) to stabilize.