(2) Connect multimeter to TI RECORDER OUTPUT (rear panel).
(3) Set TI RANGE switch to 1 mW.
(4) Position range calibrator switches as indicated in (a) through (c) below:
(a) RANGE to 1 mW.
(b) FUNCTION to CALIBRATOR.
(c) POLARITY to NORMAL.
(5) Adjust TI CAL ADJ control for a 1.000 V dc indication on multimeter.
(6) Set TI CAL FACTOR % switch to positions listed in table 9. Multimeter will
indicate within limits specified.
b. Adjustments. No adjustments can be made.
Table 9. Calibrator Factor (Models 436A, 435A, 435B, 435BOPT001, and 435BE21
Test instrument
CAL FACTOR %
(V dc)
switch settings
Min
Max
100
0.994
1.006
99
1.004
1.016
98
1.014
1.026
97
1.025
1.037
96
1.036
1.048
95
1.047
1.059
94
1.058
1.070
93
1.069
1.081
92
1.081
1.093
91
1.093
1.105
90
1.105
1.117
89
1.118
1.130
88
1.130
1.142
87
1.143
1.155
86
1.157
1.169
85
1.170
1.182
13. Calibration Factor
(Model 436A and
436AOPT9/22)
a. Performance Check
(1) Connect range calibrator to TI POWER SENSOR INPUT cable supplied
with TI.
(2) Position controls as listed in (a) through (d) below:
(a) CAL FACTOR % switch to 100.