(8) Repeat technique of (3) through (7) above for TI TIME/DIV switch settings, time
base sweep rates and oscilloscope calibrator MARKER outputs listed in table 5.
Oscilloscope will display time indications within limits specified.
b. Adjustments. No adjustments can be made.
Test instrument
TIME/DIV
calibrator
time
MARKER
switch
Time base unit
indications
settings
outputs
sweep rate
Min
Max
s
s
s
s
20
ns
0.2
0.2
1.591
1.609
s
s
s
s
50
ns
0.5
0.5
3.977
4.023
s
s
s
s
s
1
0.1
1
7.956
8.044
s
s
s
s
s
2
0.2
2
15.91
16.09
s
s
s
s
s
5
0.5
5
39.77
40.23
s
s
s
s
s
10
1
10
79.56
80.44
s
s
s
s
s
20
2
20
159.1
160.9
s
s
s
s
s
50
5
50
397.7
402.3
s
s
s
0.1 ms
0.1 ms
10
795.6
804.4
s
0.2 ms
0.2 ms
1.591 ms
1.609 ms
20
s
0.5 ms
0.5 ms
3.977 ms
4.02
ms
50
1
ms
0.1 ms
1
ms
7.956 ms
8.044 ms
2
ms
0.2 ms
2
ms
15.91
ms
16.09
ms
5
ms
0.5 ms
5
ms
39.77
ms
40.23
ms
10
ms
1
ms
10
ms
79.56
ms
80.44
ms
20
ms
2
ms
20
ms
159.1
ms
160.9
ms
50
ms
5
ms
50
ms
397.4
ms
402.6
ms
0.1 s
10
ms
0.1 s
794.9
ms
805.1
ms
0.2 s
20
ms
0.2 s
1.589 s
1.611 s
0.5 s
50
ms
0.5 s
3.974 s
4.026 s
a. Deenergize and disconnect all equipment and reinstall protective covers on TI if
necessary.
b. Annotate and affix DA label/form in accordance with TB 750-25.