TB 9-6625-2059-35
(11) Adjust resistance standard until TI multimeter indicates 30 on OHMS scale. If
resistance standard does not indicate between 28.5 and 31.5 Ω, perform b (5) below.
(12) Repeat technique of (11) above for TI multimeter RANGE switch positions and
indications listed in table 15. Resistance standard will indicate within limits specified.
Resistance standard indications (Ω)
TI multimeter
RANGE switch positions
OHMS scale indications
Min
Max
X10 Ω
30
285
315
X100 Ω
30
2850
3150
X1K Ω1
30
28.5 k
31.5 k
X10K Ω1
30
285
k
315 k
1If
required, readjust A2A1A1R123 (fig. 8) for best in-tolerance indication (R).
(13) Set TI multimeter FUNCTION switch to HI-Z and RANGE switch to 1V.
(14) Connect calibrator to TI multimeter INPUT.
(15) Adjust calibrator frequency for 1 kHz frequency and output for a 1 V indication
on TI multimeter. If calibrator does not indicate between 0.95 and 1.05 V ac, perform b (6)
below.
(16) Repeat technique of (15) above for RANGE switch settings and calibrator
frequencies listed in table 16. Calibrator will indicate within limits specified.
TI multimeter
Calibrator indications
RANGE switch settings
(V Ac)
(V)
Min
Max
.1
1 kHz
0.095
0.105
.1
20 kHz
0.095
0.105
.3
20
kHz
0.285
0.315
.3
50
Hz
0.285
.315
1
50
Hz
0.95
1.05
1
20
kHz
0.95
1.05
3
20
kHz
2.85
3.15
3
1
kHz
2.85
3.15
3
50
Hz
2.85
3.15
10
50
Hz
9.5
10.5
10
20
kHz
9.5
10.5
30
20
kHz
28.5
31.5
30
50
Hz
28.5
31.5
100
50 Hz
95
105
300
20 kHz
285
315
300
50 Hz
285
315
30