TB 9-6625-2059-35
CAUTION
Do not exceed -30 dBm into antenna input.
(4) Connect measuring receiver to signal generator output (set to minimum).
at modulation of 1 kHz at 60 percent as indicated on the measuring receiver.
(6) Connect signal generator to TI ANTENNA INPUT jack using attenuator.
(7) If multimeter does not indicate between 95 and 105 mV ac, perform b below.
b. Adjustments.
Adjust A1A13R115 (fig. 5) until
multimeter indicates 100 mV ac
(R).
a.
Performance Check
NOTE
The MM-100E multimeter will be referred to as "TI multimeter"
throughout the check.
CAUTION
Do not exceed 300 V into TI multimeter INPUT.
(1) Position TI multimeter controls as listed in (a) through (d) below:
(a) RANGE switch to .1V.
(b) FUNCTION switch to DC+.
(c) DC ZERO OFFSET control fully ccw (detent).
(d) VOL control fully ccw.
(2) Set MODULATION FREQ Hz switches to 04 000.0.
(3) Connect multimeter to EXT ACC Jack on TI. If multimeter does not indicate
zero, perform b (1) below.
(4) Set TI multimeter FUNCTION switch to DC-. If multimeter does not indicate
0, perform b (2) below.
(5) Set TI multimeter FUNCTION switch to DC+ and RANGE switch to 1V.
(6) Connect calibrator to TI multimeter INPUT using probe in table 3 (set to X1).
(7) Adjust calibrator for an indication of 10 on 0 to 10 scale. If calibrator does not
indicate between 0.97 and 1.03 V dc, perform b (3) below.
(8) Repeat technique of (7) above for TI multimeter RANGE switch settings and
meter indications listed in table 14. Calibrator will indicate within limits specified.