TB 9-6625-2069-35
(2) Press pushbuttons and enter data as listed in (a) through (i) below (both
channels for option 020).
(a)
PERIOD to 100 ns.
(b)
DELAY to 0 ns.
(c)
HIL to 1 V.
(d)
LOL to 0 V.
(e)
WIDTH to 10 ns.
(f)
LEE to 3 ns.
(g)
TRE to 3 ns.
(h)
BURST to 0.
(i)
INPUT MODE to NORM.
(3) Measure delay, between trigger output and pulse at 50 percent amplitude as
indicated on oscilloscope (fig. 2). Delay will be ≤ 1 ns.
(4) Repeat technique of (3) above, using DELAY data listed in table 6. Oscilloscope
indication will be within limits specified.
Test instrument
DELAY
indications
data
(ns)
(ns)
Min
Max
5.4
4.3
6.5
9.6
8.5
10.7
40
38.6
41.4
50
48.5
51.5
99.9 1
97.9
101.9
100
98.0
102.0
1Set
PERIOD to 1 s.
(5) Press DBL PLS pushbutton and enter data of 20 ns.
(6) Measure delay between leading edge of first pulse and leading edge of second
pulse, using standard measurement technique. Delay, as indicated on oscilloscope, will be
between 18.8 and 21.2 ns.
8