TB 9-6625-2083-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Oscilloscope Tektronix Types 453, 453A, 454, 454MOD163D, and 468. The
manufacturers' manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. TI's may vary in SN's circuit layout, adjustments, and
component values. Where applicable these variations are indicated in this bulletin. TI's
may have either BNC or UHF connectors.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported on are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Tektronix, Types 453 and 453A
Vertical deflection
Range: 5 mV/div to 10 V/div
Accuracy: 3%
Risetime
7 ns or less
5 s to 0.1 s/div except
A and B sweep
Range:
reduced to .5 s and 50 ms/div respectively
Accuracy: 3%
Calibrator
Range:
0.1 to 1 V p-p
Accuracy: 1%
Repetition
Rate:
1 kHz
Accuracy: 0.5%
Tektronix, Types 454 and 454MOD163D
Vertical deflection
Range:
5 mV/div to 10 V/div
Accuracy: 3%
Risetime
2.4 ns or less (20 mv range)
05 s/div to 5 s/div
Sweep time
Range:
Accuracy: unmag 3%, mag 4%