TB 9-6625-2083-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 995 to 1005 Hz
Fluke, Model PM6681/656
0.1%
(PM6681/656)
Accuracy:
Agilent, Model 3458A (3458A)
Range: -2.02 to + 75.2
Accuracy: 0.025%
w/HV probe Hewlett-Packard,
Model K25- 3490A
Range: -2020Vdc
Accuracy: 0.25%
Fluke, Model 5820A-5C-GHZ
Time markers:
CALIBRATOR
(5820A-5C-GHZ)
Range: 10 ns to 5 s
Accuracy: 0.75%
Voltage amplitude: 5 mV to 10 v
Range: at 1 kHz
Accuracy: 0.5%
Pulses:
Range: 1 ms to 1 s
Risetime: 1 ns or less
DVM: Range: -2020 to +75.2 V dc
Accuracy: 0.025% ( 0.5%
w/HV probe)
CALIBRATION PROCESS FOR TEKTRONIX TYPES 453 AND 453A
6. Preliminary Instructions
a. The instructions outlined in this section are preparatory to the calibration process.
Personnel should become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. When indications specified in paragraphs 8 through 16 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 16. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to TI.