TB 9-6625-2099-24
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications
AM-6785/U (Tektronix, Type 7A26)
Vertical gain
Range: 5 mV/div to 5 V/div
Accuracy: 2% with gain adjusted at 10 mV/div
Risetime
Accuracy: 7900 series, 1.8 ns (200 MHz)
w/oscilloscope
7800 series, 1.9 ns (180 MHz)
7700 series, 2.3 ns (150 MHz)
7600 series, 3.5 ns (100 MHz)
7500 series, 3.9 ns ( 90 MHz)
7400 series, 5.8 ns ( 60 MHz)
Aberrations: +4% or -4% and 6% p-p of pulse amplitude
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets, AN/GSM-286, AN/GSM-287, AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in the
calibration procedure. The following peculiar accessories are also required for this
calibration: Extender, Tektronix, Type 067-0589-00, and Standardizer 5-80 pF, APN
7916146; Oscilloscope, Tektronix, Type 7000 series; Time Base, Tektronix, Type 7B series
(compatible with TI and supplied by owner).
Table 2. Minimum Specifications of Equipment Required
Minimum use
Manufacturer and model
Common name
Specifications
(part number)
Volts out:
Fluke, Model 5820A-5C-GHZ
(5820A-5C-GHZ)
CALIBRATOR
Range: 20 mV to 20 V at 1 kHz
Accuracy: 0.5%
Square wave capabilities:
Amplitude: 30 mV to 2.5 V
Aberrations: 2%
Risetime: 0.45 ns or less