TB 9-6625-2099-35
Test
Test instrument aberrations
instrument
limits
Oscilloscope calibrator
VOLTS/DIV
EDGE
Test
Test
( 4 % and 6% pk-pk) of
switch
output
instrument
instrument
pulse amplitude
settings
aberration
display
Yes
No
(mV)
adjustments
divisions
(mV)
(MHz)
10
50
1
5
b(2)
(15) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
obtain TI display divisions listed in table 13. Measure risetime using standard risetime
techniques. Risetime will be within limits as specified in table 13 for oscilloscope series
being used.
Test
Test
Oscilloscope calibrator
instrument
Test
instrument
EDGE output
VOLTS/DIV
risetime
instrument
switch settings
limits
display
Oscilloscope
(V)
(nS)
divisions
series
Volts
(kHz)
.2
7900
1
100
5
1.8
.2
7800
1
100
5
1.9
.2
7700
1
100
5
2.2
.2
7600
1
100
5
3.5
.2
7500
1
100
5
3.9
.2
7400
1
100
5
5.8
(16) Set DISPLAY MODE switch to CH 1.
(17) Set CH 1 VOLTS/DIV switch as listed in table 14.
(18) Ensure oscilloscope calibrator CHAN 1 is connected to TI CH 1 using a 50Ω
feedthrough termination. Press oscilloscope calibrator CHANNEL pushbutton then press
blue soft pushbutton located below CHAN 1 on oscilloscope calibrator display.
(20) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
obtain oscilloscope display divisions as listed in table 14 . Measure risetime using standard
risetime techniques. Risetime will be within limits as specified in table 14 for oscilloscope
series being used.
CHANGE 2 17