TB 9-6625-2099-35
(7) Repeat technique of steps (4) through (6) above for remaining rows listed in
table 8. Oscilloscope calibrator err display indications will be within
limits specified in
Oscilloscope
Test instrument
calibrator
VOLTS/DIV
Oscilloscope calibrator
Test instrument
Err display limits
switch settings
VOLTAGE output
display divisions
(%)
5
mV
20
mV
4
2
20
mV
.1
V
5
2
50
mV
.2
V
4
2
.1
V
.5
V
5
2
.2
V
1
V
5
2
.5
V
2
V
4
2
1
V
5
V
5
2
2
V
10
V
5
2
5
V
20
V
4
2
(8) Set DISPLAY MODE switch to CH 2.
(9) Set oscilloscope calibrator output to CHAN 2 using technique in step 1 above.
(10) Set oscilloscope calibrator VOLTAGE output to 50 mV at 1 kHz.
(11) Adjust CH 2 POSITION, oscilloscope, and time base controls as necessary to
(12) Adjust CH 2 (front panel) GAIN control for 5 divisions of display.
(13) Set oscilloscope calibrator VOLTAGE output as listed in first row of table 9 .
(14) Set CH 2 VOLTS/DIV switch as listed in first row of table 9 and rotate
oscilloscope calibrator knob located below EDIT FIELD pushbutton to obtain oscilloscope
display divisions as listed in first row of table 9. Oscilloscope calibrator err display will
indicate within limits as specified in table 9; if not, perform b(5) through (12) below.
(15) Repeat technique of steps (12) through (14) above for remaining rows listed in
table 8. Oscilloscope calibrator err display indications will be within
limits specified in
Oscilloscope
calibrator
Oscilloscope calibrator
Test instrument
VOLTAGE
VOLTS/DIV
Err display limits
Test instrument
output
switch settings
(+%)
display divisions
5
mV
20
mV
4
2
20
mV
.1 V
5
2
12 CHANGE 1