TB 9-6625-2099-35
Table 9. Ch 2 Vertical Gain - Continued
Oscilloscope
calibrator
Test instrument
Oscilloscope calibrator
VOLTS/DIV
VOLTAGE
Err display limits
Test instrument
switch settings
output
display divisions
(%)
50
mV
.2
V
4
2
.1 V
.5
V
5
2
.2 V
1
V
5
2
.5 V
2
V
4
2
1
V
5
V
5
2
2
V
10
V
5
2
5
V
20
V
4
2
(16) Connect oscilloscope calibrator CHAN 1 to TI CH 1 using a 50Ω feedthrough
termination. Ensure oscilloscope calibrator CHAN 1 output is selected by using technique
in step (1) above.
(17) Set DISPLAY MODE switch to CH 1 and CH 1 VOLTS/DIV switch to 10 mV.
(18) Set oscilloscope calibrator VOLTAGE output to 60 mV at 1 kHz.
(19) Adjust CH 1 POSITION and time base controls as necessary to view display.
(20) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
perform b(13) below (CH 2, b(14) below).
(21) Remove 50Ω feedthrough termination from oscilloscope calibrator CHAN 1 and
TI CH 1. Reconnect oscilloscope calibrator CHAN 1 to TI CH 1 using a 5-80 pF
standardizer. Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton
to obtain 6 divisions 6 divisions of vertical display.
(22) Adjust standardizer for optimum square corner and flat top. If standardizer
cannot be adjusted for optimum square wave, adjust CH 1 C134 (fig. 3) (CH 1 C130 (fig. 3)
for SN B160000 and above) to midrange and readjust standardizer for optimum square
wave.